R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.5.6.2 CALCulate:MARKer:FUNCtion Subsystem<br />

The measurement window is selected by CALCulate 1 (screen A) or 2 (screen B).<br />

CALCulate Subsystem<br />

CALCulate:MARKer:FUNCtion:CENTer<br />

This command sets the center frequency of the selected measurement window equal to the frequency<br />

of the indicated marker.<br />

If marker 2, 3 or 4 is selected and used as delta marker, it is switched to the marker mode.<br />

Example: "CALC:MARK2:FUNC:CENT"<br />

'Sets the center frequency of screen A to the frequency of marker 2.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A-F<br />

This command is an "event" and therefore has no *RST value and no query.<br />

CALCulate:MARKer:FUNCtion:CSTep<br />

This command sets the step width of the center frequency in the selected measurement window to the<br />

X value of the current marker.<br />

If marker 2, 3 or 4 is selected and used as delta marker, it is switched to the marker mode.<br />

Example: "CALC2:MARK3:FUNC:CST"<br />

'Sets the center frequency of screen B to the same value as the frequency of<br />

marker 3.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A-F<br />

This command is an event and therefore has no *RST value and no query.<br />

CALCulate:MARKer:FUNCtion:DEModulation:CONTinuous ON | OFF<br />

This command switches on or off the continuous demodulation in the frequency domain (span >0) in<br />

the selected measurement window. Thus acoustic monitoring of the signals can be performed in the<br />

frequency domain. The function does not depend on the selected marker, i.e. the numeric suffix <br />

is irrelevant.<br />

Example: "CALC2:MARK3:FUNC:DEM:CONT ON"<br />

'Switches on the continuous demodulation in screen B.<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

6.54 Operating Manual 1313.9681.12 - 01