R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.17 SOURce Subsystem<br />

SOURce Subsystem<br />

The SOURce subsystem controls the output signals of the R&S <strong>FSQ</strong> if the option External Generator<br />

Control (R&S FSP-B10) is installed. The measurement window is selected by SOURce1 (screen A) and<br />

SOURce2 (screen B).<br />

6.17.1 Internal Tracking Generator<br />

SOURce:DM:STATe ON | OFF<br />

This command switches on or off the external I/Q modulation of the tracking generator in the selected<br />

measurement window.<br />

External AM and external FM are switched off, if active. This command is only valid in conjunction with<br />

the tracking generator option B9.<br />

Example: "SOUR2:DM:STAT ON"<br />

'Switches on the external I/Q modulation of the tracking generator for screen B.<br />

Characteristics: *RST value: OFF<br />

SCPI: conform<br />

Mode: A<br />

SOURce:FM:DEViation 100Hz to 10MHz<br />

This command defines the maximum frequency deviation at 1 V input voltage at the FM input of the<br />

tracking generator.<br />

The possible value range is 100 Hz to 10 MHz in steps of one decade.<br />

This command is only available in connection with the Tracking Generator Option B9.<br />

Example: "SOUR:FM:DEV 1MHz"<br />

'Sets the maximum frequency deviation of the tracking generator for screen A to 1<br />

MHz.<br />

Characteristics: *RST value: 100 Hz<br />

SCPI: conform<br />

Mode: A<br />

SOURce:FM:STATe ON | OFF<br />

This command switches on or off the external frequency modulation of the tracking generator in the<br />

selected measurement window.<br />

External I/Q modulation is switched off, if active. This command is only valid in conjunction with the<br />

tracking generator option B9.<br />

Example: "SOUR:FM:STAT ON"<br />

'Switches on the external frequency modulation of the tracking generator for<br />

screen A.<br />

Characteristics: *RST value: OFF<br />

SCPI: conform<br />

Mode: A<br />

6.212 Operating Manual 1313.9681.12 - 01