R&S <strong>FSQ</strong> Instrument Functions<br />

NUMBER OF<br />

SWEEPS<br />

<strong>Analyzer</strong> Mode<br />

The measured values can be reset by switching the AVERAGE ON / OFF softkey off<br />

and on again.<br />

Remote command: CALC:MARK:FUNC:SUMM:AVER ON<br />

CALC:MARK:FUNC:SUMM:PPE:AVER:RES?<br />

CALC:MARK:FUNC:SUMM:RMS:AVER:RES?<br />

CALC:MARK:FUNC:SUMM:MEAN:AVER:RES?<br />

CALC:MARK:FUNC:SUMM:SDEV:AVER:RES?<br />

The NUMBER OF SWEEPS softkey activates the entry of the number of sweeps for<br />

maximum or average value calculation.<br />

SINGLE SWEEP mode The R&S <strong>FSQ</strong> performs sweeps until the<br />

selected number of sweeps is reached and<br />

stops then.<br />

CONTINUOUS SWEEP mode Averaging is carried out until the selected<br />

number of sweeps is reached. After that,<br />

averaging is performed in continuous mode<br />

and is then continued as running averaging.<br />

Calculation of the maximum peak (MAX<br />

HOLD) is performed continuously irrespective<br />

of the selected number of sweeps.<br />

The valid range values is 0 to 32767.<br />

Depending on the specified number of sweeps, averaging is carried out according to<br />

the following rules:<br />

NUMBER OF SWEEPS = 0 Continuous averaging is carried out over 10<br />

measured values.<br />

NUMBER OF SWEEPS = 1 No averaging, maxhold or minhold is carried<br />

out.<br />

NUMBER OF SWEEPS > 1 Averaging is carried out over the set number of<br />

measured values.<br />

This setting is equivalent to the setting of the sweep count in the TRACE menu.<br />

Remote command: SWE:COUN <br />

Example<br />

The mean power of a GSM burst with 0 dBm nominal power at 800 MHz is to be<br />

measured.<br />

[PRESET] Set the R&S <strong>FSQ</strong> to the default setting.<br />

[FREQ: CENTER: 800 MHz] Set the center frequency to 800 MHz.<br />

[SPAN: ZERO SPAN] Select time domain display (span = 0 Hz).<br />

[AMPT: 0 dBm] Set the reference level to 0 dBm.<br />

[BW: RES BW MANUAL: 30<br />

kHz]<br />

Set the resolution bandwidth to 30 kHz in line<br />

with the requirements of the GSM standard.<br />

4.84 Operating Manual 1313.9681.12 - 01