R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

Example: "CALC:MARK:FUNC:SUMM:MODE REL"<br />

'Switches the time domain power measurement to relative.<br />

Characteristics: *RST value: ABSolute<br />

SCPI: device-specific<br />

Mode: A-T<br />

CALCulate:MARKer:FUNCtion:SUMMary:PHOLd ON | OFF<br />

CALCulate Subsystem<br />

This command switches on or off the peak-hold function for the active time domain power<br />

measurement in the indicated measurement window. The function is independent of the marker<br />

selection, i.e. the numeric suffix of :MARKer is irrelevant. It is only available in the time domain<br />

(span = 0). The peak-hold function is reset by switching it off and on again.<br />

Example: "CALC:MARK:FUNC:SUMM:PHOL ON"<br />

'Switches on the function in screen A.<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A-T<br />

The peak-hold function is reset by switching off and on, again.<br />

CALCulate:MARKer:FUNCtion:SUMMary:PPEak:AVERage:RESult?<br />

This command is used to query the result of the measurement of the averaged positive peak value in<br />

the selected measurement window. The query is only possible if averaging has been activated<br />

previously using CALCulate:MARKer:FUNCtion:SUMMary:AVERage.<br />

The function is independent of the marker selection, i.e. the numeric suffix in MARKer is<br />

irrelevant. It is only available in the time domain (span = 0).<br />

A complete sweep with synchronization to sweep end must be performed between switching on the<br />

function and querying the measured value to obtain a valid query result. This is only possible in singlesweep<br />

mode.<br />

Example: "INIT:CONT OFF"<br />

'Switches to single-sweep mode.<br />

"CALC:MARK:FUNC:SUMM:PPE ON"<br />

'Switches on the function in screen A.<br />

"CALC:MARK:FUNC:SUMM:AVER ON"<br />

'Switches on the calculation of average in screen A.<br />

"INIT;*WAI"<br />

'Starts a sweep and waits for the end.<br />

"CALC:MARK:FUNC:SUMM:PPE:AVER:RES?"<br />

'Outputs the result of screen A.<br />

Characteristics: *RST value: -<br />