R&S <strong>FSQ</strong> Instrument Functions<br />

Basic Settings<br />

FULL SCREEN The FULL SCREEN softkey selects the display of one diagram. This corresponds to<br />

the default setting of R&S <strong>FSQ</strong>.<br />

In the FULL SCREEN mode, it is possible to switch between two different device<br />

settings by selecting the active window (screen A or screen B).<br />

Switching between SCREEN A and SCREEN B is performed by means of the corresponding<br />

key in the hotkey bar (for details refer to “Mode Selection – Hotkey Bar” on<br />

page 4.7).<br />

It should be noted that the measurements in the FULL SCREEN mode are performed<br />

only in the visible (active) window.<br />

SPLIT<br />

SCREEN<br />

REF LEVEL<br />

COUPLED<br />

CENTER B =<br />

MARKER A |<br />

CENTER A =<br />

MARKER B<br />

The active window is marked by A or B on the right of the diagram.<br />

Remote command: DISP:FORM SING<br />

DISP:WIND:SEL<br />

The SPLIT SCREEN softkey selects the display of two diagrams. The upper diagram<br />

is designated SCREEN A, the lower diagram SCREEN B.<br />

Switching between SCREEN A and SCREEN B is performed via the corresponding<br />

key in the hotkey bar. The active window is marked by highlighting fields A and<br />

B<br />

on the right of the diagram.<br />

Remote command: DISP:FORM SPL<br />

The REF LEVEL COUPLED softkey switches the coupling of the reference level on<br />

and off. In addition to the reference level, the mixer level and input attenuation are<br />

coupled with one another.<br />

For the level measurement, the same reference level and input attenuation must be<br />

set for the two diagrams.<br />

Remote command: INST:COUP RLEV<br />

The CENTER B = MARKER A and CENTER A = MARKER B softkeys couple the<br />

center frequency in diagram B with the frequency of marker 1 in diagram A and the<br />

center frequency in diagram B with the frequency of marker 1 in diagram B. The two<br />

softkeys are mutually exclusive.<br />

This coupling is useful, e.g. for viewing the signal at the marker position in diagram<br />

A with higher frequency resolution or in the time domain in diagram B.<br />

If marker 1 is off, it is switched on and set to the maximum of the trace in the active<br />

diagram.<br />

Remote command: INST:COUP CF_B<br />

INST:COUP CF_A<br />