18.12.2012 Views

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

R&S <strong>FSQ</strong> Instrument Functions<br />

<strong>Analyzer</strong> Mode<br />

FULL SPAN The FULL SPAN softkey sets the span to the full frequency range of R&S <strong>FSQ</strong>.<br />

Remote command: FREQ:SPAN:FULL<br />

ZERO SPAN The ZERO SPAN softkey sets the span to 0 Hz. The x-axis becomes the time axis<br />

with the grid lines corresponding to 1/10 of the current sweep time (SWT).<br />

Remote command: FREQ:SPAN 0Hz<br />

LAST SPAN After changing the span setting the LAST SPAN softkey activates the previous setting.<br />

With this function a fast change between overview measurement (FULL SPAN)<br />

and detailed measurement (manually set center frequency and span) is possible.<br />

FREQ AXIS<br />

(LIN LOG)<br />

AMPT<br />

Only values > 0 Hz are restored, i.e. a transition between time and frequency<br />

domain is not possible.<br />

Remote command: --<br />

The FREQ AXIS (LIN LOG) softkey switches between linear and logarithmic scaling<br />

of the frequency axis. Switch over is only possible if the stop/start frequency ratio is<br />

≥1.4.<br />

The default state is LIN.<br />

The logarithmic frequency axis is only available in analyzer mode and it is not available<br />

in zero span mode, in external mixer mode, with frequency offset or if the ratio<br />

stop frequency / start frequency is below 1.4.<br />

Remote command: DISP:WIND:TRAC:X:SPAC LIN<br />

4.5.3 Level Display Setting and RF Input Configuration – AMPT Key<br />

The AMPT key is used to set the reference level, the maximum level and the display<br />

range of the active window as well as the input impedance and the input attenuation<br />

of the RF input.<br />

The AMPT key opens a menu for setting the reference level and the input attenuation<br />

of the active window. The data entry for the reference level (REF LEVEL softkey)<br />

is opened automatically.<br />

Further settings regarding level display and attenuation can be made in this menu.<br />

REF LEVEL<br />

RANGE LOG 100 dB<br />

RANGE LOG MANUAL<br />

RANGE LINEAR � RANGE LINEAR %<br />

RANGE LINEAR dB<br />

UNIT � dBm / dBmV / dBµV /<br />

dBµΑ / dBµW / VOLT /<br />

AMPERE / WATT<br />

RF INPUT (AC DC)<br />

4.14 Operating Manual 1313.9681.12 - 01

Hooray! Your file is uploaded and ready to be published.