R&S <strong>FSQ</strong> Instrument Functions<br />

<strong>Analyzer</strong> Mode<br />

= MARKER The = MARKER softkey sets the step size coupling to MANUAL and the step size to<br />

the value of the marker. This function is especially useful during measurements of<br />

the signal harmonic content at the marker position because by entering the center<br />

frequency each stroke of the STEP key selects the center frequency of another harmonic.<br />

Remote command: --<br />

MANUAL The MANUAL softkey activates the window for entering a fixed step size.<br />

Remote command: FREQ:CENT:STEP 120MHz<br />

Softkeys in time domain:<br />

0.1 * RBW The 0.1 * RBW softkey sets the step size for the center frequency entry to 10% of<br />

the resolution bandwidth.<br />

AUTO 0.1 * RBW corresponds to the default setting.<br />

Remote command: FREQ:CENT:STEP:LINK RBW<br />

FREQ:CENT:STEP:LINK:FACT 10PCT<br />

0.5 * RBW The 0.5 * RBW softkey sets the step size for the center frequency entry to 50% of<br />

the resolution bandwidth.<br />

Remote command: FREQ:CENT:STEP:LINK RBW<br />

FREQ:CENT:STEP:LINK:FACT 50PCT<br />

X * RBW The X * RBW softkey allows the factor defining the center frequency step size to be<br />

entered as % of the resolution bandwidth.<br />

Values between 1 and 100% in steps of 1% are allowed. The default setting is 10%.<br />

Remote command: FREQ:CENT:STEP:LINK RBW<br />

FREQ:CENT:STEP:LINK:FACT 20PCT<br />

= CENTER The = CENTER softkey sets the step size coupling to MANUAL and the step size to<br />

the value of the center frequency. This function is especially useful during measurements<br />

of the signal harmonic content because by entering the center frequency<br />

each stroke of the STEP key selects the center frequency of another harmonic.<br />

Remote command: --<br />

= MARKER The = MARKER softkey sets the step size coupling to MANUAL and the step size to<br />

the value of the marker. This function is especially useful during measurements of<br />

the signal harmonic content at the marker position because by entering the center<br />

frequency each stroke of the STEP key selects the center frequency of another harmonic.<br />

Remote command: --<br />

MANUAL The MANUAL softkey activates the window for entering a fixed step size.<br />

Remote command: FREQ:CENT:STEP 120MHz<br />

START The START softkey activates the window for manually entering the start frequency.<br />

Operating Manual 1313.9681.12 - 01 4.11