R&S <strong>FSQ</strong> Instrument Functions<br />

MEAS<br />

<strong>Analyzer</strong> Mode<br />

The current marker search limit settings (LEFT LIMIT, RIGHT LIMIT, THRESHOLD,<br />

EXCLUDE LO) are taken into account.<br />

Remote command: CALC:MARK:MAX:AUTO ON | OFF<br />

CALC:MARK:MIN:AUTO ON | OFF<br />

4.5.13 Power Measurements – MEAS Key<br />

With its power measurement functions the R&S <strong>FSQ</strong> is able to measure all the necessary<br />

parameters with high accuracy in a wide dynamic range.<br />

A modulated carrier is almost always used (except e.g. SSB-AM) for high-frequency<br />

transmission of information. Due to the information modulated upon the carrier, the<br />

latter covers a spectrum which is defined by the modulation, the transmission data<br />

rate and the signal filtering. Within a transmission band each carrier is assigned a<br />

channel taking into account these parameters. In order to ensure error-free transmission,<br />

each transmitter must be conforming to the specified parameters. These<br />

include among others:<br />

• the output power,<br />

• the occupied bandwidth, i.e. the bandwidth which must contain a defined<br />

percentage of the power and<br />

• the power dissipation allowed in the adjacent channels.<br />

Additionally the menu contains functions to determine the modulation depth of AM<br />

modulated signals and to measure the 3rd order intercept point.<br />

The measurements and the corresponding settings are selected in the MEAS menu.<br />

TIME DOM POWER �<br />

CHAN PWR ACP MULT CARR ACP �<br />

OCCUPIED BANDWIDTH �<br />

SIGNAL STATISTIC �<br />

C/N / C/NO �<br />

MODULATION DEPTH<br />

IQ MODE<br />

SPECTRUM EMISSION MASK �<br />

SPURIOUS EMISSIONS �<br />

SELECT MARKER<br />

Side menu<br />

TOI<br />

TOI MKR CALC SRCH<br />

HARMONIC DISTOR �<br />

The MEAS key opens the menu to select and set the power measurement.<br />

The following measurements can be selected:<br />

• Power in the time domain (“TIME DOM POWER” on page 4.80)<br />

Operating Manual 1313.9681.12 - 01 4.79