R&S <strong>FSQ</strong> Instrument Functions<br />

DELAY COMP<br />

(ON OFF)<br />

<strong>Analyzer</strong> Mode<br />

Fig. 4.6 Setting GATE DELAY and GATE LENGTH in time domain by means of lines GD and GL<br />

On quitting the GATE SETTINGS menu, R&S <strong>FSQ</strong> returns to the previous screen.<br />

The DELAY COMP (ON OFF) softkey enables the filter group delay compensation<br />

for the external trigger and IF power trigger. If a bursted signal is analyzed in zero<br />

span and the delay compensation is on, a change of the RBW will not change the<br />

rising slope of the signal.<br />

Default is OFF.<br />

Remote command: TRIG:HOLD:ADJ:AUTO ON|OFF<br />

4.5.7 Selection and Setting of Traces – TRACE Key<br />

The R&S <strong>FSQ</strong> is capable of displaying up to three different traces at a time in a diagram.<br />

A trace consists of a maximum of 625 pixels on the horizontal axis (frequency<br />

or time). If more measured values than pixels are available, several measured values<br />

are combined in one pixel.<br />

The traces are selected using the SELECT TRACE softkey in the menu of the<br />

TRACE key.<br />

The traces can individually be activated for a measurement or frozen after completion<br />

of a measurement. Traces that are not activated are blanked.<br />

The display mode can be selected for each trace. Traces can be overwritten in each<br />

measurement (CLEAR/WRITE mode), averaged over several measurements<br />

(AVERAGE mode), or a maximum or minimum value can be determined from several<br />

measurements and displayed (MAX HOLD or MIN HOLD).<br />

Individual detectors can be selected for the various traces. The autopeak detector<br />

displays maximum and minimum values connected by a vertical line. The max peak<br />

detector and min peak detector display the maximum and minimum value of the<br />

level within a pixel. The sample detector displays the instantaneous value of the<br />

level at a pixel. The rms detector displays the power (rms value) of the measured<br />

values within a pixel, the average detector the average value.<br />