R&S <strong>FSQ</strong> Instrument Functions<br />

TRD FACTOR<br />

NAME<br />

TRD FACTOR<br />

UNIT<br />

TRD FACTOR<br />

VALUES<br />

Basic Settings<br />

During editing, a transducer factor remains stored in the background until the factor<br />

edited is saved with the SAVE TRD FACTOR softkey or until the table is closed. A<br />

factor that was edited by mistake can be restored by leaving the entry function.<br />

Name - Entry of name<br />

The TRD FACTOR NAME softkey is used to enter the name of the transducer factor.<br />

A maximum of 8 characters is permissible for the name. The characters have to<br />

comply with the convention of DOS file names. If the name exceeds 8 characters,<br />

the name is truncated. The unit automatically stores all transducer factors with the<br />

extension .TDF. If an existing name is changed, the factor stored under the previous<br />

name is retained and will not be overwritten automatically with the new name. The<br />

previous factor can be deleted at a later time, using DELETE. Thus, factors can be<br />

copied.<br />

Remote command: CORR:TRAN:SEL <br />

Unit – Selection of unit<br />

The TRD FACTOR UNIT softkey gives access to a box for selecting the unit of the<br />

transducer factor.<br />

The default setting is dB.<br />

Remote command: CORR:TRAN:UNIT <br />

Entry of value<br />

The TRD FACTOR VALUES softkey sets the scrollbar on the first reference value.<br />

The desired reference values for FREQUENCY and TDF/dB must be entered in<br />

ascending order of frequencies. After the frequency has been entered, the scrollbar<br />

automatically goes to the associated level value.<br />

The table can be edited after entry of the first value using the INSERT LINE and<br />

DELETE LINE softkeys. To change individual values later on, the value has to be<br />

selected and a new one entered.<br />

Remote command: CORR:TRAN:DATA ,<br />

Interpolation - Selection of interpolation<br />

Linear or logarithmic interpolation can be performed between the frequency reference<br />

values of the table. The ENTER key allows the user to select LIN or LOG (toggle<br />

function).<br />

Remote command: CORR:TRAN:SCAL LIN|LOG<br />