R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.22 UNIT Subsystem<br />

The UNIT subsystem is used to switch the base unit of setting parameters.<br />

The numeric suffix at UNIT selects the measurement window.<br />

UNIT Subsystem<br />

UNIT:POWer DBM | DBPW | WATT | DBUV | DBMV | VOLT | DBUA | AMPere | V | A | W | DB |<br />

DBPT | PCT | UNITLESS | DBUV_MHZ | DBMV_MHZ | DBUA_MHZ | DBUV_M | DBUA_M |<br />

DBUV_MMHZ | DBUA_MMHZ<br />

This command selects the unit for power for the selected measurement window.<br />

DBxx_MHZ indicates the units dBxx/MHz and DBxx_MMHZ indicates the units dBxx/mMHz (the<br />

measured level is referenced to a 1 MHz bandwidth).<br />

Example: "UNIT:POW DBM"<br />

'Sets the power unit for screen A to dBm.<br />

Characteristics: *RST value: DBM<br />

SCPI: conform<br />

Mode: A<br />

Operating Manual 1313.9681.12 - 01 6.255