18.12.2012 Views

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.16.7 SENSe:FREQuency Subsystem<br />

SENSe Subsystem<br />

The SENSe:FREQuency subsystem defines the frequency axis of the active display. The frequency axis<br />

can either be defined via the start/stop frequency or via the center frequency and span.<br />

The measurement windows are selected by SENSe1 (screen A) and SENSe2 (screen B).<br />

[SENSe:]FREQuency:CENTer 0 to fmax<br />

This command defines the center frequency of the R&S <strong>FSQ</strong> or the measuring frequency for span = 0.<br />

Example: "FREQ:CENT 100MHz"<br />

Characteristics: *RST value: fmax /2 with fmax = maximum frequency<br />

SCPI: conform<br />

Mode: A<br />

[SENSe:]FREQuency:CENTer:STEP 0 to fmax<br />

This command defines the step width of the center frequency.<br />

Example: "FREQ:CENT:STEP 120MHz"<br />

Characteristics: *RST value: - (AUTO 0.1 × SPAN is switched on)<br />

SCPI: conform<br />

Mode: A<br />

[SENSe:]FREQuency:CENTer:STEP:LINK SPAN | RBW | OFF<br />

This command couples the step width of the center frequency to span (span >0) or to the resolution<br />

bandwidth (span = 0) or cancels the couplings.<br />

Parameter: SPAN = Coupling to frequency display range (for span > 0)<br />

RBW = Coupling to resolution bandwidth (for span = 0)<br />

OFF = manual input, no coupling.<br />

Example: "FREQ:CENT:STEP:LINK SPAN"<br />

Characteristics: *RST value: SPAN<br />

SCPI: device-specific<br />

Mode: A<br />

[SENSe:]FREQuency:CENTer:STEP:LINK:FACTor 1 to 100 PCT<br />

This command couples the step width of the center frequency with a factor to the span (span >0) or to<br />

the resolution bandwidth (span = 0).<br />

Example: "FREQ:CENT:STEP:LINK:FACT 20PCT"<br />

Characteristics: *RST value: - (AUTO 0.1 × SPAN is switched on)<br />

SCPI: device-specific<br />

Mode: A<br />

6.172 Operating Manual 1313.9681.12 - 01