18.12.2012 Views

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

R&S <strong>FSQ</strong> Instrument Functions<br />

CHANNEL<br />

SPACING<br />

<strong>Analyzer</strong> Mode<br />

Measurements in the time domain (FAST ACP ON) are performed in the zero span<br />

mode. The channel limits are indicated by vertical lines. For measurements requiring<br />

channel bandwidths deviating from those defined in the selected standard the IBW<br />

method is to be used.<br />

Refer to section “Filter Types” on page 4.26 for a list of available filters.<br />

When measuring according to the IBW method (FAST ACP OFF) the bandwidths of<br />

the different adjacent channels are to be entered numerically. Since all adjacent<br />

channels often have the same bandwidth, the other channels Alt1 and Alt2 are set to<br />

the bandwidth of the adjacent channel on entering the adjacent-channel bandwidth<br />

(ADJ). Thus only one value needs to be entered in case of equal adjacent channel<br />

bandwidths. The same holds true for the ALT2 channels (alternate channels 2) when<br />

the bandwidth of the ALT1 channel (alternate channel 1) is entered.<br />

The channel spacings can be set separately by overwriting the table from top to<br />

bottom.<br />

The entry "TX" is only available for the multi carrier ACP measurement.<br />

Remote command: SENS:POW:ACH:BWID:CHAN 14kHz<br />

SENS:POW:ACH:BWID:ACH 1kHz<br />

SENS:POW:ACH:BWID:ALT1 14kHz<br />

SENS:POW:ACH:BWID:ALT2 14kHz<br />

The CHANNEL SPACING softkey opens a table for defining the channel spacings<br />

for the TX channels as well as for the adjacent channels.<br />

The entry "TX" is only available for the multi carrier ACP measurement.<br />

TX channels<br />

The spacing between every TX channels can be defined separately. Therefore a TX<br />

spacing 1-2 for the spacing between the first and the second carrier, a TX spacing 2-<br />

3 for the spacing between the second and the third carrier and so on can be defined.<br />

In order to allow a convenient setup for the system with equal TX channel spacing,<br />

the value of TX spacing 1-2 will be copied in all the spacing below after entry, the TX<br />

spacing 2-3 will be copied in all the spacing below after entry and so forth.<br />

For different spacings, a setup from top to bottom is necessary.<br />

If the spacings are not equal, the channel distribution according to the center<br />

frequency is as follows:<br />

• Odd number of TX channels:<br />

The middle TX channel is centered to center frequency.<br />

Operating Manual 1313.9681.12 - 01 4.97