R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.16.9 SENSe:LIST Subsystem<br />

SENSe Subsystem<br />

The commands of this subsystem are used for measuring the power at a list of frequency points with different<br />

device settings. The measurement is always performed in the time domain (span = 0 Hz).<br />

A new trigger event is required for each test point (exception: Trigger FREE RUN).<br />

The results are output as a list in the order of the entered frequency points. The number of results per<br />

test point depends on the number of concurrently active measurements (peak/RMS/average).<br />

Selection of concurrently active measurements and setting of parameters that are constant for the whole<br />

measurement is via a configuration command (SENSe:LIST:POWer:SET). This also includes the setting<br />

for trigger and gate parameters.<br />

The following setting parameters can be selected independently for each frequency point:<br />

• <strong>Analyzer</strong> frequency<br />

• Reference level<br />

• RF attenuation of attenuator (only with option B25)<br />

• Resolution filter<br />

• Resolution bandwidth<br />

• Video bandwidth<br />

• Measurement time<br />

• Detector<br />

The number of frequencies is limited to 100 entries.<br />

The commands of this subsystem can be used in two different ways:<br />

1. Instrument setup, measurement and querying of the results in a single command:<br />

With this method, there is the least delay between the measurement and the result output. However,<br />

it requires the control computer to wait for the response from the instrument.<br />

2. Instrument setup and querying of the result list at the end of the measurement:<br />

With this method, the control computer may be used for other activities while the measurement is being<br />

performed. However, more time is needed for synchronization via service request.<br />

[SENSe:]LIST:POWer:RESult?<br />

This command queries the result of a previous list measurement as configured and initiated with<br />

SENSe:LIST:POWer[:SEQuence]. The measured results are output in a list of floating point values<br />

separated by commas. The unit of the results depends on the setting made with the CALC:UNIT<br />

command.<br />

This command may be used to obtain measurement results in an asynchronous way, using the service<br />

request mechanism for synchronization with the end of the measurement.<br />

If no measurement results are available, the command will return a query error.<br />

Example:<br />

Settings that are not directly included in commands of this subsystem can be configured<br />

by sending the corresponding commands prior to the SENSe:LIST commands.<br />

Please note that changes to the trigger level have to be executed in time domain<br />

(span = 0 Hz) in order to take effect for the SENSe:LIST commands.<br />

6.176 Operating Manual 1313.9681.12 - 01