R&S <strong>FSQ</strong> Getting Started<br />

Noise Measurements<br />

Fig. 2.15 Approximating the channel filter by sweeping with a small resolution bandwidth<br />

The following steps are performed:<br />

• The linear power of all the trace pixels within the channel is calculated.<br />

Pi = 10(Li/10)<br />

where<br />

Pi = power of the trace pixel i<br />

Li = displayed level of trace point i<br />

• The powers of all trace pixels within the channel are summed up and the sum is<br />

divided by the number of trace pixels in the channel.<br />

• The result is multiplied by the quotient of the selected channel bandwidth and the<br />

noise bandwidth of the resolution filter (RBW).<br />

Since the power calculation is performed by integrating the trace within the channel<br />

bandwidth, this method is also called the IBW method (Integration Bandwidth<br />

method).<br />

Bandwidth selection (RBW)<br />

For channel power measurements, the resolution bandwidth (RBW) must be small<br />

compared to the channel bandwidth, so that the channel bandwidth can be defined<br />

precisely. If the resolution bandwidth which has been selected is too wide, this may<br />

have a negative effect on the selectivity of the simulated channel filter and result in<br />

the power in the adjacent channel being added to the power in the transmit channel.<br />

A resolution bandwidth equal to 1% to 3% of the channel bandwidth should, therefore,<br />

be selected. If the resolution bandwidth is too small, the required sweep time<br />

becomes too long and the measurement time increases considerably.<br />

Detector selection<br />

Resolution filter<br />

Sweep<br />

Channel bandwith<br />

Since the power of the trace is measured within the channel bandwidth, only the<br />

sample detector and RMS detector can be used. These detectors provide measured<br />

values that make it possible to calculate the real power. The peak detectors (Pos<br />

Peak, Neg Peak and Auto Peak) are not suitable for noise power measurements as<br />

no correlation can be established between the peak value of the video voltage and<br />

power.<br />

With the sample detector, a value (sample) of the IF envelope voltage is displayed<br />

at each trace pixel. Since the frequency spans are very large compared with the resolution<br />

bandwidth (span/RBW >500), sinewave signals present in the noise might be<br />

2.22 Operating Manual 1313.9681.12 - 01<br />

-3 dB