R&S <strong>FSQ</strong> Instrument Functions<br />

EXT SRC ON /<br />

OFF<br />

SELECT<br />

GENERATOR<br />

External Generator Control – Option R&S FSP-B10<br />

The EXT SRC ON / OFF softkey switches the external generator on or off.<br />

It can only be switched on successfully if the generator has been selected by means<br />

of SELECT GENERATOR and configured correctly by means of FREQUENCY<br />

SWEEP. If one of these conditions is not fulfilled, an error message will be output.<br />

When switching on the external generator by means of EXT SRC ON, the<br />

R&S <strong>FSQ</strong> switches off the internal tracking generator and starts programming the<br />

generator settings via the GPIB interface IEC2.<br />

Programming requires takeover of the remote control at this interface by the<br />

R&S <strong>FSQ</strong>. To avoid any access conflicts, ensure that no other controller is connected<br />

to the IEC2 interface or the external generator when selecting EXT SRC<br />

ON.<br />

The maximum stop frequency is limited to the maximum generator frequency. This<br />

upper limit is automatically reduced by the set frequency offset of the generator<br />

and a set multiplication factor.<br />

With the external generator switched on, the FFT filters (FILTER TYPE FFT in the<br />

menu BW) are not available.<br />

If an error occurs on the GPIB when programming the external generator, the generator<br />

will automatically be switched off and the following error message will be<br />

output:<br />

When the external generator is switched off using EXT SRC OFF, the GPIB control<br />

is handed over again at the IEC2 interface, i.e. a different controller will then take<br />

over the control of the signal generator.<br />

Remote command: SOUR:EXT ON<br />

The SELECT GENERATOR softkey opens a table for selection of the generator and<br />

definition of GPIB address and control interface.<br />

The table permits configuration of two generators so that switching between two different<br />

configurations is easily possible.<br />

The individual fields contain the following settings:<br />

• SRC<br />

Index of generator selected<br />