R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

CALCulate:DELTamarker:MINimum:RIGHt<br />

CALCulate Subsystem<br />

This command positions the specified delta marker to the next higher minimum value to the right of the<br />

current value (i.e. ascending X values). The corresponding delta marker will be activated first, if<br />

necessary.<br />

Example: "CALC2:DELT:MIN:RIGH"<br />

'Sets delta marker 1 in screen B to the next higher minimum value to the right of<br />

the current value.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is an event and therefore has no *RST value and no query.<br />

CALCulate:DELTamarker:MODE ABSolute | RELative<br />

This command switches between relative and absolute frequency input of the delta marker (or time<br />

with span = 0). It affects all delta markers independent of the measurement window.<br />

Example: "CALC:DELT:MODE ABS"<br />

'Switches the frequency/time indication for all delta markers to absolute values.<br />

"CALC:DELT:MODE REL"<br />

'Switches the frequency/time indication for all delta markers to relative to marker 1.<br />

Characteristics: *RST value: REL<br />

SCPI: device-specific<br />

Mode: A<br />

Mode:<br />

CALCulate:DELTamarker[:STATe] ON | OFF<br />

This command switches on and off the delta marker if delta marker 1 is selected. If marker 2, 3 or 4 is<br />

selected and used as a marker, it is switched to delta marker mode. If the corresponding marker is not<br />

activated, it will be activated and positioned on the maximum of the measurement curve.<br />

If no numeric suffix is indicated, delta marker 1 is selected automatically.<br />

The measurement windows are selected via CALCulate1 (screen A) or 2 (screen B).<br />

Example: "CALC:DELT3 ON"<br />

'Switches marker 3 in screen A to delta marker mode.<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

CALCulate:DELTamarker:TRACe 1 to 3<br />

This command assigns the selected delta marker to the indicated measurement curve. The selected<br />

measurement curve must be active, i.e. its state must be different from "BLANK".<br />

The measurement windows are selected via CALCulate1 (screen A) or 2 (screen B).<br />

Example: "CALC:DELT3:TRAC 2"<br />

'Assigns deltamarker 3 to trace 2 in screen A.<br />

Operating Manual 1313.9681.12 - 01 6.19