R&S <strong>FSQ</strong> Instrument Functions<br />

SELECT USER<br />

STD<br />

SAVE AS USER<br />

STD<br />

DELETE USER<br />

STD<br />

<strong>Analyzer</strong> Mode<br />

The softkey SELECT USER STD selects and activates a user defined ACP standard.<br />

After selecting the standard, the user defined atndard becomes available for selection<br />

in the ACP standard list (USER list item). The corresponding filename is shown<br />

in brackets.<br />

Remote command: CALC:MARK:FUNC:POW:PRES <br />

As of firmware version 4.4x, you can save the current ACP configuration in a file and<br />

use it as a user defined ACP standard. The data set is stored on the instruments<br />

harddisk in XML file format.<br />

Softkey SAVE AS USER STD opens a dialog box to store the user defined configuration.<br />

The following list shows the parameters that you can customize:<br />

• Number of Adjacent Channels<br />

• Channel Bandwidth of TX, ADJ and ALT channels<br />

• Channel Spacings<br />

• Resolution Bandwidth, Video Bandwidth<br />

• ACP Limit State and ACP Limits<br />

• Sweeptime, Sweeptime Coupling<br />

• Detector<br />

• Trace Mode (Clr/Write,..)<br />

Note that the ACP User Standard is not supported for Fast ACP and Multi Carrier<br />

ACP Measurements.<br />

Remote command: use SENSe:POWer Subsystem to configure a<br />

user standard<br />

CALC:MARK:FUNC:POW:STAN:SAVE <br />

CALC:MARK:FUNC:POW:STAN:CAT?<br />

Opens a dialog box to delete user defined ACP standards.<br />

Remote command: CALC:MARK:FUNC:POW:STAN:DEL <br />