R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

Mode: A-T<br />

SCPI: device-specific<br />

TRACe:IQ:SRATe 400 Hz to 100 MHz<br />

TRACe Subsystem<br />

This command sets the sampling rate for the I/Q data acquisition. Thus the sample rate can be<br />

modified without affecting the other settings.<br />

Return value: 400 Hz to 100 MHz (326.4 MHz with the option R&S <strong>FSQ</strong>-B72)<br />

sampling rates >81.6 MHz require a bandwidth of 50 MHz, set with the TRAC:IQ:<br />

SET command.<br />

Example: TRAC:IQ:SRAT 4MHZ<br />

Characteristics: *RST value: 32 MHz<br />

SCPI: device-specific<br />

Mode: A-T<br />

TRACe:IQ[:STATe] ON | OFF<br />

For using these default settings with command TRAC:IQ:DATA? the following<br />

minimum buffer sizes for the response data are recommended:<br />

ASCII format: 10 kBytes<br />

Binary format: 2 kBytes<br />

If = CFILter, the sample rate is determined by the selected filter bandwidth.<br />

In this case, the currently set sample rate can be determined using the<br />

query command. Entering a value causes an execution error if =<br />

CFILter.<br />

This command switches the I/Q data acquisition on or off.<br />

It is possible to combine I/Q data measurement with gated trigger function for following<br />

settings:<br />

• Sample rate 81.6 MHz<br />

• Gate source EXTERN<br />

The gate parameters (gate mode, gate polarity, gate delay, and gate length) are<br />

configured by the subsystem SENSe:SWEep:EGATe, when TRACE:IQ:STATe is<br />

set to ON.<br />

All other measurement functions will be switched off as soon as the I/Q measurement<br />

function is switched on. Additionally, a trace display is not possible in this<br />

operating mode. Therefore all traces are set to "BLANK". Finally, split screen operation<br />

will automatically be stopped.<br />

Example: TRAC:IQ ON<br />

'Switches on I/Q data acquisition<br />

TRAC:IQ:ONL ON<br />

'Switches the online output of the I/Q data on.<br />