R&S <strong>FSQ</strong> Instrument Functions<br />

REFERENCE<br />

FREQUENCY<br />

REF BW (NORM<br />

WIDE)<br />

NOISE SRC<br />

(ON OFF)<br />

SIGNAL<br />

SOURCE<br />

Basic Settings<br />

Once the R&S <strong>FSQ</strong> has switched to the internal reference signal it stays with the<br />

internal reference. You have to switch back to the external reference signal manually<br />

(press REFERENCE EXTERNAL and then activate REFERENCE EXT (INT) again).<br />

The current source (internal or external) is indicated in the lower part of the softkey.<br />

Remote command: ROSC:SOUR EAUT<br />

The REFERENCE FREQUENCY softkey sets the frequency of the external reference.<br />

The range is from 1 MHz to 20 MHz with the default value being 10 MHz.<br />

The reference frequency does not change after a preset to maintain the setup of a<br />

test system.<br />

Remote command: ROSC:EXT:FREQ 10 MHZ<br />

The REF BW (NORM WIDE) softkey controls the PLL bandwidth used to sznchronize<br />

with the external reference signal. Possible settings are NORM (PLL bandwidth<br />

about 3 Hz) or WIDE (PLL bandwidth about 30 Hz). The default setting is NORM.<br />

The reference settings do not change after a preset to maintain the setup of a test<br />

system.<br />

The softkey is available only if the external reference frequency is 10 MHz.<br />

This function requires a synthersizer board revision with hardware code 7 or above .<br />

You can check this information in the “HWC” column of the hardware info window<br />

(see “HARDWARE INFO” on page 4.198).<br />

Remote command: ROSC:EXT:PLL NORM | WIDE<br />

4.6.3.2 External Noise Source<br />

The NOISE SRC (ON OFF) softkey switches on or off the supply voltage for an<br />

external noise source which is connected to the NOISE SOURCE connector on the<br />

rear panel of the instrument.<br />

Remote command: DIAG:SERV:NSO ON<br />

4.6.3.3 <strong>Signal</strong> Source<br />

The SIGNAL SOURCE softkey opens the submenu for switching on or off the YIG<br />

FILTER, the IQ DITHER and RF PATH.<br />

YIG FILTER (ON OFF)<br />

IQ DITHER ON OFF<br />

RF PATH<br />

BASEBAND ANALOG �<br />

BASEBAND DIGITAL � DIGITAL IN FULL SCALE<br />

DIGITAL IN SAMPLE RATE<br />

FULL SCALE AUTO SET /<br />

SAMPLE RATE AUTO SET<br />

DIGITAL BB INFO<br />

EX-IQ BOX<br />