R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.5.3 CALCulate:ESPectrum Subsystem<br />

CALCulate Subsystem<br />

The following chapter describes remote control commands that configure the list evaluation for the Spectrum<br />

Emission Mask measurement.<br />

CALCulate:ESPectrum:PSEarch | PEAKsearch:AUTO ON | OFF<br />

This command activates or deactivates the list evaluation of the Spectrum Emission Mask<br />

measurement.<br />

Example: “CALC:ESP:PSE:AUTO OFF”<br />

’Deactivates the list evaluation<br />

Characteristics: RST value: ON<br />

SCPI: device-specific<br />

Mode: A<br />

CALCulate:ESPectrum:PSEarch | PEAKsearch:MARGin -200dB … +200dB<br />

This command sets the margin used for the limit check/peak search of the Spectrum Emission Mask<br />

measurement.<br />

Example: “CALC:ESP:PSE:MARG 10"<br />

' sets the margin to 10 dB<br />

Characteristics: *RST value: 6 dB<br />

SCPI: device-specific<br />

Mode: A<br />