R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.16.13 SENSe:ROSCillator Subsystem<br />

SENSe Subsystem<br />

This subsystem controls the reference oscillator. The numeric suffix in SENSe is irrelevant for the commands<br />

of this subsystem.<br />

[SENSe:]ROSCillator:EXTernal:FREQuency 1MHz to 20MHz<br />

This command informs the instrument on the frequency of the external reference oscillator. This<br />

frequency is used as starting point for the synchronization of the internal reference frequencies.<br />

The value of the external reference frequency (1 MHz to 20 MHz) is rounded in steps of 1 Hz.<br />

Example: "ROSC:EXT:FREQ 5MHz"<br />

Characteristics: *RST value: -<br />

SCPI: conform<br />

Mode: A<br />

*RST is not influencing this parameter.<br />

[SENSe:]ROSCillator:EXTernal:PLL NORMal | WIDE<br />

This command controls the PLL bandwidth used to synchronize with the external reference signal and<br />

is available only with an external reference frequency of 10 MHz.<br />

The reference settings are not changed if a preset occurs to maintain the specific setup of a test<br />

system. The factory default is NORMal.<br />

NORMal Uses a PLL bandwidth of about 3 Hz<br />

WIDE Uses a PLL bandwidth of about 30 Hz<br />

This function requires a synthesizer board revision with hardware code 7 or above. You can check this<br />

information in the “HWC” column of the hardware info window (see “HARDWARE INFO” on<br />

page 4.198).<br />

Example: "ROSC:EXT:PLL NORM"<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

[SENSe:]ROSCillator[:INTernal]:TUNe 0 to 4095<br />

This command defines the value for the tuning of the internal reference oscillator.<br />

The reference oscillator should be tuned only if an error has been detected in the frequency accuracy<br />

check. After rebooting the instrument, the factory-set reference frequency or the previously saved<br />

reference frequency is restored.<br />

This command is only available at service level 1.<br />

Example: "ROSC:TUN 128"<br />

Characteristics: *RST value: –<br />

SCPI: device-specific<br />

Mode: A<br />

6.204 Operating Manual 1313.9681.12 - 01