R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

DISPlay Subsystem<br />

The trace is switched off (= BLANK in manual operation) with the DISPlay[:WINDow]:<br />

TRACe[:STATe] command.<br />

The number of measurements for AVERage, MAXHold and MINHold is defined with the [SENSe:<br />

]AVERage:COUNt or [SENSe:]SWEep:COUNt command.<br />

Synchronization to the end of the indicated number of measurements is only possible in single-sweep<br />

mode.<br />

If calculation of average values is active, selection between logarithmic and linear averaging is<br />

possible. For more detail see command [SENSe:]AVERage:TYPE.<br />

Example: "INIT:CONT OFF"<br />

'Switching to single-sweep mode.<br />

"SWE:COUN 16"<br />

'Sets the number of measurements to 16.<br />

"DISP:WIND1:TRAC3:MODE MAXH"<br />

'Switches on the calculation of the for trace 3 in screen A.<br />

"INIT;*WAI"<br />

'Starts the measurement and waits for the end of the 16 sweeps.<br />

Characteristics: *RST value: WRITe for TRACe1, STATe OFF for TRACe2/3<br />

SCPI: device-specific<br />

Mode: A<br />

DISPlay[:WINDow]:TRACe:MODE:HCONtinuous ON | OFF<br />

This command specifies whether or not the traces with peak or minimum value detection are reset after<br />

specific parameter changes.<br />

Usually the measurement must be restarted after a parameter change, before an evaluation of the<br />

measurement results is performed (e.g. with a marker). In cases in which a change causes a<br />

compulsory new measurement, the trace is automatically reset in order to prevent erroneous<br />

measurements of previous measurement results (e.g. when the span changes). For applications in<br />

which this behavior is not desired, this mechanism can be switched off.<br />

Parameter: OFF: The traces are reset after specific parameter changes.<br />

ON: The reset mechanism is switched off.<br />

Example: "DISP:WIND1:TRAC3:MODE:HCON ON"<br />

'The reset mechanism is switched off for measurement window 1.<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

DISPlay[:WINDow]:TRACe[:STATe] ON | OFF<br />

This command switches on or off the display of the corresponding trace in the selected measurement<br />

window.<br />

Example: "DISP:WIND1:TRAC3 ON"<br />

Characteristics: *RST value: ON for TRACe1, OFF for TRACe2 and 3<br />

SCPI: conform<br />

Mode: A<br />