R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.9 FORMat Subsystem<br />

FORMat Subsystem<br />

The FORMat subsystem specifies the data format of the data transmitted from and to the instrument.<br />

FORMat[:DATA] ASCii | REAL | UINT [, 8 | 32]<br />

This command specifies the data format for the data transmitted from the instrument to the control PC.<br />

The format settings are valid for the binary transmission of trace data (see also TRACe[:DATA]).<br />

Example: "FORM REAL,32"<br />

"FORM ASC"<br />

"FORM UINT,8"<br />

Characteristics: *RST value: ASCii<br />

SCPI: conform<br />

Mode: A<br />

The data format is either ASCII or one of the formats REAL. ASCII data are transmitted in plain text,<br />

separated by commas. REAL data are transmitted as 32-bit IEEE 754 floating-point numbers in the<br />

"definite length block format".<br />

The FORMat command is valid for the transmission of trace data. The data format of trace data<br />

received by the instrument is automatically recognized, regardless of the format which is programmed.<br />

FORMat:DEXPort:DSEParator POINt | COMMA<br />

This command defines which decimal separator (decimal point or comma) is to be used for outputting<br />

measurement data to the file in ASCII format. Different languages of evaluation programs (e.g.<br />

Microsoft Excel) can thus be supported.<br />

Example: "FORM:DEXP:DSEP POIN<br />

'Sets the decimal point as separator.<br />

Characteristics: *RST value: -- (factory setting is POINt; *RST does not affect setting)<br />

SCPI: device-specific<br />

Mode: A<br />

Incorrect format setting will result in numerical conversion, which may lead to<br />

incorrect results.<br />