R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

"CALC:MARK:FUNC:SUMM:RMS:AVER:RES?"<br />

'Outputs the result of screen A.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A-T<br />

This command is only a query and therefore has no *RST value.<br />

CALCulate:MARKer:FUNCtion:SUMMary:RMS:PHOLd:RESult?<br />

CALCulate Subsystem<br />

This command queries the result of the measurement of the RMS value with active peak hold in the<br />

selected measurement window. The query is only possible only if the peak hold function has been<br />

activated previously using CALCulate:MARKer:FUNCtion:SUMMary:PHOLd.<br />

The function is independent of the marker selection, i.e. the numeric suffix of :MARKer is<br />

irrelevant. It is only available in the time domain (span = 0).<br />

A complete sweep with synchronization to sweep end must be performed between switching on the<br />

function and querying the measured value to obtain a valid query result. This is only possible in singlesweep<br />

mode.<br />

Example: "INIT:CONT OFF"<br />

'Switches to single-sweep mode.<br />

"CALC:MARK:FUNC:SUMM:RMS ON"<br />

'Switches on the function in screen A.<br />

"CALC:MARK:FUNC:SUMM:PHOL ON"<br />

'Switches on the peak value measurement in screen A.<br />

"INIT;*WAI"<br />

'Starts a sweep and waits for the end.<br />

"CALC:MARK:FUNC:SUMM:RMS:PHOL:RES?"<br />

'Outputs the result of screen A.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A-T<br />

This command is only a query and therefore has no *RST value.<br />

CALCulate:MARKer:FUNCtion:SUMMary:RMS:RESult?<br />

This command queries the result of the measurement of the RMS power value in the selected<br />

measurement window.<br />

The function is independent of the marker selection, i.e. the numeric suffix of :MARKer is<br />

irrelevant. It is only available in the time domain (span = 0).<br />

A complete sweep with synchronization to sweep end must be performed between switching on the<br />

function and querying the measured value to obtain a valid query result. This is only possible in singlesweep<br />

mode.<br />

Example: "INIT:CONT OFF"<br />

'Switches to single-sweep mode.<br />

"CALC:MARK:FUNC:SUMM:RMS ON"<br />

'Switches on the function in screen A.<br />

Operating Manual 1313.9681.12 - 01 6.87