18.12.2012 Views

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

Characteristics: *RST value: --<br />

SCPI: device-specific<br />

Mode: A<br />

This command is an event and therefore has no *RST value and no query.<br />

SYSTem:LXI:MDEScription<br />

SYSTem Subsystem<br />

This command shows the LXI instrument description. Also, the instrument description can be changed<br />

using this command.<br />

Example: "SYST:LXI:INFO?"<br />

’Queries the current description<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

SYSTem:LXI:PASSword <br />

This command shows or changes the LXI password. The default password is 'LxiWebIfc'.<br />

Example: "SYST:LXI:PASS '123456'"<br />

’Sets a new password (123456).<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

SYSTem:MSIZe? MBOard | B100<br />

This command outputs the memory size of the mainboard or of the R&S <strong>FSQ</strong>-B100 option.<br />

Parameter:<br />

MBOard Mainboard<br />

B100 R&S <strong>FSQ</strong>-B100<br />

Example: "SYST:MSIZ? MBO"<br />

Outputs the memory size of the mainboard.<br />

Characteristics: *RST value: --<br />

SCPI: device-specific<br />

Mode: A<br />

SYSTem:PASSword[:CENable] 'password'<br />

This command enables access to the service functions by means of the password.<br />

Example: "SYST:PASS 'XXXX'"<br />

Characteristics: *RST value: –<br />

SCPI: conform<br />

Mode: A<br />

This command is an event and therefore has no *RST value and no query.<br />

6.236 Operating Manual 1313.9681.12 - 01

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!