R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

CALCulate:MARKer:COUNt:FREQuency?<br />

CALCulate Subsystem<br />

This command queries the result of the frequency counter for marker 1 in the selected measurement<br />

window. Before the command, the frequency counter should be switched on and a complete<br />

measurement performed to obtain a valid count result. Therefore, a single sweep with synchronization<br />

must be performed between switching on the frequency counter and querying the count result.<br />

Example: "INIT:CONT OFF"<br />

'Switches to single-sweep mode.<br />

"CALC:MARK ON"<br />

'Switches the marker in screen A on.<br />

"CALC:MARK:COUN ON"<br />

'Switches the frequency counter for the marker.<br />

"INIT;*WAI"<br />

'Starts a sweep and waits for the end.<br />

"CALC:MARK:COUN:FREQ?"<br />

'Outputs the measured value of the marker in screen A.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is only a query and therefore has no *RST value.<br />

CALCulate:MARKer:COUNt:RESolution 0.1 | 1 | 10 | 100 | 1000 | 10000 Hz<br />

This command specifies the resolution of the frequency counter in the selected measurement window.<br />

The setting is independent of the selected marker, i.e. the numeric suffix in MARKer is<br />

irrelevant.<br />

Example: "CALC:MARK:COUN:RES 1kHz"<br />

'Sets the resolution of the frequency counter to 1 kHz.<br />

Characteristics: *RST value: 1kHz<br />

SCPI: device-specific<br />

Mode: A<br />

CALCulate:MARKer:LOEXclude ON | OFF<br />

This command switches the local oscillator suppression for peak search on or off. This setting is valid<br />

for all markers and delta markers in all measurement windows. The numeric suffixes 1|2 and 1...4 are<br />

irrelevant.<br />

Example: "CALC:MARK:LOEX ON"<br />

Characteristics: *RST value: ON<br />

SCPI: device-specific<br />

Mode: A-F<br />

CALCulate:MARKer:MAXimum:AUTO ON | OFF<br />

This command switches an automatic maximum peak search for marker 1 at the end of each particular<br />

sweep on and off. The current marker search limit settings (LEFT LIMIT, RIGHT LIMIT, THRESHOLD,<br />

EXCLUDE LO) are taken into account. The numeric suffix at MARKer is irrelevant.<br />

6.46 Operating Manual 1313.9681.12 - 01