R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

SENSe Subsystem<br />

'Configuration of the status reporting system for the generation of an SRQ<br />

on operation complete<br />

*ESE 1<br />

*SRE 32<br />

'Configuring and starting the measurement<br />

"SENSe:LIST:POWer 935.2MHz,-20dBm,10dB,OFF,NORM,1MHz,3MHz,434us,0,<br />

935.4MHz,-20dBm,10dB,10dB,NORM,30kHz,100kHz,434us,0,<br />

935.6MHz,-20dBm,10dB,20dB,NORM,30kHz,100kHz,434us,0;<br />

*OPC"<br />

'Further actions of the control computer during measurement<br />

to<br />

'Response to service request<br />

On SRQ:<br />

SENSe:LIST:POWer:RESult?<br />

Characteristics: *RST value: --<br />

SCPI: device-specific<br />

Mode: A<br />

[SENSe:]LIST:POWer[:SEQuence] ,,,, ,<br />

, , ,<br />

This command configures the list of settings (max. 100 entries) for the multiple power measurement<br />

and starts a measurement sequence. When synchronizing the command with *OPC, a service request<br />

is generated as soon as all frequency points are processed and the defined number of individual<br />

measurements is reached.<br />

To reduce the setting time, all indicated parameters are set up simultaneously at each test point.<br />

The query form of the command processes the list and immediately returns the list of results. The<br />

number of results per test point depends on the setting of the "SENSe:LIST:POWer:SET" command.<br />

The following parameters are the settings for an individual frequency point. They<br />

are repeated for every other frequency point.<br />

Parameter:<br />

: Receive frequency for the signal to be measured (= center frequency in manual<br />

operation)<br />

Range of values: 0 Hz to max. frequency, depending on the instrument<br />

model.<br />

: Reference level<br />

Range of values: +30 dBm to -70 dBm in 10 dB steps<br />

+30 dBm to -75 dBm in 5 dB steps with El. Attenuator<br />

option B25<br />

: RF input attenuation<br />

Range of values: 0 dB to 70 dB in 10 dB steps<br />

Operating Manual 1313.9681.12 - 01 6.177