R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

CALCulate:LIMit:UPPer:THReshold <br />

CALCulate Subsystem<br />

This command defines an absolute threshold value for limit lines with relative y-axis scaling<br />

independently of the measurement window. The absolute threshold value is used in the limit check as<br />

soon as it exceeds the relative limit value.<br />

The unit must correspond to the unit selected with CALC:LIM:UNIT (except dB which is not possible).<br />

If no unit is indicated, the unit defined with CALC:LIM:UNIT is automatically used (exception: dBm<br />

instead of dB).<br />

In analyzer mode, the units DEG, RAD, S, HZ, PCT are not available.<br />

Example: "CALC:LIM2:UPP:THR -35DBM"<br />

'Defines an absolute threshold value for limit line 2.<br />

Characteristics: *RST value: -200 dBm<br />

SCPI: device-specific<br />

Mode: A<br />

6.44 Operating Manual 1313.9681.12 - 01