R&S <strong>FSQ</strong> Instrument Functions<br />

SOURCE ON/<br />

OFF<br />

SOURCE<br />

POWER<br />

POWER<br />

OFFSET<br />

MODULATION OFF<br />

POWER SWEEP POWER SWP ON/OFF<br />

START POWER<br />

STOP POWER<br />

Tracking Generator – Option R&S FSU-B9<br />

Additional softkeys are available in the displayed menus for controlling an external<br />

generator if option External Generator Control R&S FSP-B10 is fitted. For detailed<br />

information see section “External Generator Control – Option R&S FSP-B10” on<br />

page 4.235.<br />

The SOURCE ON/OFF softkey switches the tracking generator on or off. Default<br />

setting is OFF<br />

• When the tracking generator is switched on the maximum stop frequency is<br />

limited to 3.6 GHz. This upper limit is automatically reduced by a frequency<br />

offset set up for the tracking generator.<br />

•<br />

• In order to meet the data sheet accuracy for measurements with active tracing<br />

generator the start frequency must be set to ≥ 3 × Resolution Bandwidth.<br />

• The minimum sweeptime for measurements with data sheet accuracy is 100 ms<br />

in frequency domain (span > 0 Hz). Selecting a sweeptime below this limit will<br />

result in the sweeptime indicator field SWT being supplied with a red asterisk<br />

and the message UNCAL being displayed.<br />

• FFT filters (FILTER TYPE FFT in BW menu) are not available when the tracking<br />

generator is active.<br />

Remote command: OUTP:STAT ON<br />

The SOURCE POWER softkey allows the tracking generator output power to be<br />

selected.<br />

The output power can be set in 0.1 dB steps from -30 dBm to +5 dBm (-100 to + 5<br />

dBm with option R&S FSU-B12).<br />

If the tracking generator is off, it is automatically switched on when an output power<br />

value is entered.<br />

The default output power is -20 dBm.<br />

Remote command: SOUR:POW -20dBm<br />

The POWER OFFSET softkey allows selection of a constant level offset for the<br />

tracking generator.<br />

With this offset for example attenuators or amplifiers at the output connector of the<br />

tracking generator can be taken into account for the displayed output power values<br />

on screen or during data entry.<br />

4.222 Operating Manual 1313.9681.12 - 01