R&S <strong>FSQ</strong> Remote Control – Basics<br />

5.4.3.2 Return to Manual Operation<br />

Starting Remote Control<br />

Return to manual operation is possible via the front panel or via RS-232 interface.<br />

Manually:<br />

Press the LOCAL softkey or the PRESET key.<br />

• Before the transition, command processing must be completed as otherwise<br />

transition to remote control is performed immediately<br />

• The keys can be enabled again by sending the control string "@LOC" via RS-<br />

232 (see chapter “Maintenance and Instrument Interfaces”, section “RS-232-C<br />

Interface (COM)” on page 8.7).<br />

Via RS-232:<br />

v24puts(port,"@LOC");<br />

Set instrument to manual operation.<br />

5.4.3.3 Restrictions<br />

The following restrictions apply if the unit is remote-controlled via the RS-232-C<br />

interface:<br />

No interface messages, only control strings (see interface description in chapter<br />

“Maintenance and Instrument Interfaces”, section “RS-232-C Interface (COM)” on<br />

page 8.7).<br />

Only the Common Commands *OPC? can be used for command synchronization,<br />

*WAI and *OPC are not available.<br />

Block data cannot be transmitted.<br />

5.4.4 Remote Control in a Network (LAN Interface)<br />

Via the LAN interface, the R&S <strong>FSQ</strong> can be remotely controlled in a local area network.<br />

The LAN interface consists of a connector, a network interface card and protocols<br />

(VXI-11 and RSIB). For details on the connector and its use refer to the Quick Start<br />

Guide, chapter 1, "Front and Rear Panel".<br />

Instrument access via VXI11 or RSIB is usually achieved from high level programming<br />

platforms by using VISA as an intermediate abstraction layer. VISA encapsulates<br />

the low level VXI, RSIB or even GPIB function calls and thus makes the<br />

transport interface transparent for the user. The necessary VISA library is available<br />

as a separate product. For details contact your local R&S sales representative.<br />

Operating Manual 1313.9681.12 - 01 5.7