R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

CALCulate:DELTamarker:LINK ON | OFF<br />

CALCulate Subsystem<br />

This command links delta marker 1 to marker 1. If you change the horizontal position of the marker, so<br />

does the delta marker.<br />

The suffix at DELTamarker can only be 1 or not present, because the functionality is only available for<br />

marker 1 and delta marker 1.<br />

Example: "CALC1:DELT1:LINK ON"<br />

Switches the link of marker1/delta marker 1 on.<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

CALCulate:DELTamarker:MAXimum:LEFT<br />

This command positions the specified delta marker to the next smaller maximum value to the left of<br />

the current value (i.e. descending X values). The corresponding delta marker will be activated first, if<br />

necessary.<br />

Example: "CALC:DELT:MAX:LEFT"<br />

'Sets delta marker 1 in screen A to the next smaller maximum value to the left of<br />

the current value.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is an event and therefore has no *RST value and no query.<br />

CALCulate:DELTamarker:MAXimum:NEXT<br />

This command positions the specified delta marker to the next smaller maximum value of the trace.<br />

The corresponding delta marker will be activated first, if necessary.<br />

Example: "CALC1:DELT2:MAX:NEXT"<br />

'Sets delta marker 2 in screen A to the next smaller maximum value.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is an event and therefore has no *RST value and no query.<br />

CALCulate:DELTamarker:MAXimum[:PEAK]<br />

This command positions the specified delta marker to the current maximum value of the trace. If<br />

necessary, the corresponding delta marker will be activated first.<br />

Example: "CALC2:DELT3:MAX"<br />

'Sets delta marker 3 in screen B to the maximum value of the associated trace.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is an event and therefore has no *RST value and no query.<br />