R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

CALCulate Subsystem<br />

CALCulate:LIMit:ACPower:ALTernate:ABSolute -200DBM to 200DBM,<br />

-200DBM to 200DBM<br />

This command defines the absolute limit value for the lower/upper alternate adjacent channel power<br />

measurement (Adjacent Channel Power) in the selected measurement window.<br />

The numeric suffix after ALTernate denotes the alternate channel. The numeric suffixes<br />

are irrelevant for this command.<br />

It should be noted that the absolute limit value for the limit check has no effect as soon as it is below<br />

the relative limit value defined with CALCulate:LIMit:ACPower:ALTernate:<br />

RELative. This mechanism allows automatic checking of the absolute basic values defined in mobile<br />

radio standards for the power in adjacent channels.<br />

Parameter: The first value is the limit for the lower and the upper alternate adjacent channel.<br />

The second limit value is ignored but must be indicated for reasons of compatibility<br />

with the FSE family.<br />

Example: "CALC:LIM:ACP:ALT2:ABS -35DBM, -35DBM"<br />

'Sets the absolute limit value in screen A for the power in the lower and upper<br />

second alternate adjacent channel to -35 dBm.<br />

Characteristics: *RST value: -200DBM<br />

SCPI: device-specific<br />

Mode: A<br />

CALCulate:LIMit:ACPower:ALTernate:ABSolute:STATe ON | OFF<br />

This command activates the limit check for the alternate adjacent channels in the selected<br />

measurement window for adjacent channel power measurement (Adjacent Channel Power).<br />

Before the command, the limit check must be globally switched on for the channel/adjacent channel<br />

power with the command CALCulate:LIMit:ACPower:STATe ON.<br />

The numeric suffix after ALTernate denotes the alternate channel. The numeric suffixes<br />

are irrelevant for this command.<br />

The result can be queried with CALCulate:LIMit:ACPower:ALTernate:RESult?. It<br />

should be noted that a complete measurement must be performed between switching on the limit<br />

check and the result query, since otherwise no valid results are available.<br />

Example: "CALC:LIM:ACP:ALT2 30DB, 30DB"<br />

'Sets the relative limit value in screen A for the power in the lower and upper<br />

second alternate adjacent channel to 30 dB below the channel power.<br />

"CALC:LIM:ACP:ALT2:ABS -35DBM, -35DBM"<br />

'Sets the absolute limit value in screen A for the power in the lower and upper<br />

second alternate adjacent channel to -35 dBm.<br />

"CALC:LIM:ACP ON"<br />

'Switches on globally the limit check for the channel/adjacent channel<br />

measurement in screen A.<br />

"CALC:LIM:ACP:ACH:REL:STAT ON"<br />

Switches on the check of the relative limit values for the alternative adjacent<br />

channels in screen A.<br />

"CALC:LIM:ACP:ACH:ABS:STAT ON"<br />

'Switches on the check of absolute limit values for the alternative adjacent<br />

channels in screen A.<br />