R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

CALCulate:MARKer:FUNCtion:TOI[:STATe] ON | OFF<br />

CALCulate Subsystem<br />

This command initiates the measurement of the third-order intercept point.<br />

A two-tone signal with equal carrier levels is expected at the RF input of the instrument. Marker 1 and<br />

marker 2 (both normal markers) are set to the maximum of the two signals. Delta marker 3 and delta<br />

marker 4 are positioned to the intermodulation products. The delta markers can be modified separately<br />

afterwards with the commands CALCulate:DELTamarker3:X and CALCulate:DELTamarker4:<br />

X.<br />

The third-order intercept is calculated from the level spacing between the normal markers and the delta<br />

markers.<br />

The numeric suffix of :MARKer is irrelevant for this command.<br />

Example: "CALC:MARK:FUNC:TOI ON"<br />

'Switches on the measurement of the third-order intercept in screen A.<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

CALCulate:MARKer:FUNCtion:ZOOM <br />

This command defines the range to be zoomed around marker 1 in the selected measurement window.<br />

Marker 1 is activated first, if necessary.<br />

The subsequent frequency sweep is stopped at the marker position and the frequency of the signal is<br />

counted. This frequency becomes the new center frequency, and the zoomed span is set. In order to<br />

recognize the end of the operation the synchronization to the sweep end should be activated. This is<br />

only possible in single-sweep mode.<br />

Example: "INIT:CONT OFF"<br />

'Switches to single-sweep mode<br />

"CALC:MARK:FUNC:ZOOM 1kHz;*WAI"<br />

'Activates zooming in screen A and waits for its end.<br />

for phase noise mode<br />

"CALC1:MARK1:FUNC:ZOOM 10"<br />

'Zooms in around marker 1 by a factor of 10.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is an event and therefore has no *RST value and no query.<br />

6.62 Operating Manual 1313.9681.12 - 01