R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

CALCulate:STATistics:SCALe:AUTO ONCE<br />

+<br />

CALCulate Subsystem<br />

This command optimizes the level setting of the instrument depending on the measured peak power,<br />

in order to obtain maximum instrument sensitivity.<br />

To obtain maximum resolution, the level range is set as a function of the measured spacing between<br />

peak power and the minimum power for the APD measurement and of the spacing between peak<br />

power and mean power for the CCDF measurement. In addition, the probability scale for the number<br />

of test points is adapted.<br />

Example: "CALC:STAT:SCAL:AUTO ONCE;*WAI"<br />

'Adapts the level setting for statistical measurements and activates the<br />

synchronization.<br />

Characteristics: *RST value: --<br />

SCPI: device-specific<br />

Mode: A<br />

This command is an event and therefore has no *RST value and no query.<br />

CALCulate:STATistics:SCALe:X:MPOSition 0 … 100<br />

This command defines the relative X position of the mean power value for the CCDF measurement.<br />

The default position is 0% (left corner of the grid). This function is only available if the statistics<br />

measurement function CCDF is switched on.<br />

Example: "CALC:STAT:SCAL:X:MPOS 10.0"<br />

' set the mean power position to 10%<br />

Characteristics: *RST value: 0<br />

SCPI: device-specific<br />

Mode: A<br />

CALCulate:STATistics:SCALe:X:RANGe 1dB to 200dB<br />

This command defines the level range for the x-axis of the measurement diagram. The setting is<br />

identical to the level range setting defined with the command DISPlay:WINDow:TRACe:Y:SCALe.<br />

Example: "CALC:STAT:SCAL:X:RANG 20dB"<br />

Characteristics: *RST value: 100dB<br />

SCPI: device-specific<br />

Mode: A<br />

Subsequent commands have to be synchronized with *WAI, *OPC or *OPC? to the<br />

end of the auto range process which would otherwise be aborted.<br />

6.96 Operating Manual 1313.9681.12 - 01