R&S <strong>FSQ</strong> Instrument Functions<br />

<strong>Analyzer</strong> Mode<br />

The level of the intermodulation products depends on the level of the useful signals.<br />

If the two useful signals are increased by 1 dB, the level of the intermodulation products<br />

increases by 3 dB, which means that spacing aD3 between intermodulation signals<br />

and useful signals is reduced by 2 dB. This is illustrated in Fig. 4.15<br />

Dependence of intermodulation level on useful signal level.<br />

Output<br />

level<br />

Compression<br />

Useful signal<br />

1<br />

1 1<br />

Intermodulation<br />

product<br />

Intercept point<br />

Fig. 4.15 Dependence of intermodulation level on useful signal level<br />

Input level<br />

The useful signals at the two port output increase proportionally with the input level<br />

as long as the two port is in the linear range. A level change of 1 dB at the input<br />

causes a level change of 1 dB at the output. Beyond a certain input level, the two<br />

port goes into compression and the output level stops increasing. The intermodulation<br />

products of the third order increase three times as much as the useful signals.<br />

The intercept point is the fictitious level where the two lines intersect. It cannot be<br />

measured directly since the useful level is previously limited by the maximum two<br />

port output power.<br />

It can be calculated from the known line slopes and the measured spacing aD3 at a<br />

given level according to the following formula.<br />

IP3 =<br />

aD3 -------- + P<br />

2 N<br />

The 3 rd order intercept point (TOI), for example, is calculated for an intermodulation<br />

of 60 dB and an input level PU of -20 dBm according to the following formula:<br />

60<br />

IP3 – ----- +<br />

( – 20dBm)<br />

– 10dBm<br />

2<br />

TOI The TOI softkey enables the measurement of the 3rd order intercept point.<br />

A two-tone signal with equal carrier levels is expected at the R&S <strong>FSQ</strong> input.<br />

MARKER 1 and MARKER 2 (both normal markers) are set to the maximum of the<br />

two signals. MARKER 3 and MARKER 4 (both delta markers) are placed on the<br />

intermodulation products. When the function is enabled, the frequency entry is activated<br />

for the delta markers. They can be set manually.<br />

Operating Manual 1313.9681.12 - 01 4.123<br />

3<br />

(16)<br />

(17)