18.12.2012 Views

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

R&S <strong>FSQ</strong> Error Messages<br />

Error code Error text in the case of queue poll<br />

Error explanation<br />

SCPI-Specific Error Messages<br />

-158 String data not allowed<br />

The command contains a valid string at a position which is not allowed.<br />

Example: A text parameter is set in quotation marks, INPut:COUPling "DC"<br />

-160 Block data error<br />

The command contains faulty block data.<br />

-161 Invalid block data<br />

The command contains faulty block data.<br />

Example: An END message was received prior to reception of the expected number<br />

of data.<br />

-168 Block data not allowed<br />

The command contains valid block data at an invalid position.<br />

Example: The command *RCL requires a number to be indicated.<br />

-170 Expression error<br />

The command contains an invalid mathematical expression.<br />

-171 Invalid expression<br />

The command contains an invalid mathematical expression.<br />

Example: The expression contains mismatching parentheses.<br />

-178 Expression data not allowed<br />

The command contains a mathematical expression at an invalid position.<br />

Error code Error text in the case of queue poll<br />

Error explanation<br />

-200 Execution error<br />

Error on execution of the command.<br />

-201 Invalid while in local<br />

The command is not executable while the device is in local due to a hard local control.<br />

Example: The device receives a command which would change the rotary knob state,<br />

but the device is in local so the command can not be executed.<br />

-202 Settings lost due to rtl<br />

A setting associated with hard local control was lost when the device changed to<br />

LOCS from REMS or to LWLS from RWLS.<br />

-210 Trigger error<br />

Error on triggering the device.<br />

-211 Trigger ignored<br />

The trigger (GET, *TRG or trigger signal) was ignored because of device timing<br />

considerations.<br />

Example: The device was not ready to respond.<br />

-212 Arm ignored<br />

An arming signal was ignored by the device.<br />

-213 Init ignored<br />

Measurement initialization was ignored as another measurement was already in<br />

progress.<br />

Operating Manual 1313.9681.12 - 01 9.5

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!