R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.12 INPut Subsystem<br />

INPut Subsystem<br />

The INPut subsystem controls the input characteristics of the RF inputs of the instrument. The measurement<br />

windows are assigned to INPut1 (screen A) and INPut2 (screen B).<br />

INPut:ATTenuation 0 to 75dB<br />

This command programs the input attenuator. To protect the input mixer against damage from<br />

overloads, the setting 0 dB can be obtained by entering numerals, not by using the command DEC.<br />

The step width is 5 dB, the range is 0 dB to..75 dB.<br />

In analyzer mode, the attenuation set on the step attenuator is coupled to the reference level of the<br />

instrument. If the attenuation is programmed directly, the coupling to the reference level is switched off.<br />

Example: "INP:ATT 40dB"<br />

'Sets the attenuation on the attenuator to 40 dB and switches off the coupling to the<br />

reference level.<br />

Characteristics: *RST value: 10 dB (AUTO is set to ON)<br />

SCPI: conform<br />

Mode: A<br />

INPut:ATTenuation:AUTO ON | OFF<br />

This command automatically couples the input attenuation to the reference level (state ON) or<br />

switches the input attenuation to manual entry (state OFF). The minimum input attenuation set with the<br />

coupling switched on is 10 dB (with electronic attenuator option: 5 dB).<br />

Example: "INP:ATT:AUTO ON"<br />

'Couples the attenuation set on the attenuator to the reference level.<br />

Characteristics: *RST value: ON<br />

SCPI: conform<br />

Mode: A<br />

INPut:ATTenuation:PROTection:RESet<br />

The R&S <strong>FSQ</strong> is equipped with an overload protection mechanism. This mechanism becomes active<br />

as soon as the power at the input mixer exceeds a value of 27 dBm. It ensures that the connection<br />

between RF input and input mixer is cut off.<br />

The command resets the attenuator into the state that it had before the overload condition was<br />

detected. It re-connects the RF input with the input mixer.<br />

Example: "INP:ATT:PROT:RES"<br />

Characteristics: *RST value: --<br />

SCPI: device-specific<br />

Mode: A<br />

This command comes into effect only if the reason for the overload condition has<br />

been eliminated. Otherwise the connection between RF input and input mixer is<br />

left open.<br />

This command is an event and therefore has no query and no *RST value.<br />

6.124 Operating Manual 1313.9681.12 - 01