R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

[SENSe:]LIST:STANdard:CATalog?<br />

SENSe Subsystem<br />

This command queries the names of customized spurious emission measurement configuration files<br />

currently stored on the hard disk of the R&S <strong>FSQ</strong>.<br />

The command queries the files that are in the folder "D:\R_S\INSTR\SPURS_STD". The file extension<br />

(*.xml) is not returned.<br />

The numeric suffix at SENSe is irrelevant.<br />

This command is a query and therefore has no *RST value.<br />

Example: "LIST:STAN:CAT?"<br />

'Returns the name of spurious emission configuration files.<br />

Characteristics: *RST value: --<br />

SCPI: device-specific<br />

Mode: A<br />

[SENSe:]LIST:STANdard:DELete <br />

This command deletes a file that contains a spurious emission measurement configuration.<br />

The default path to the file is "D:\R_S\INSTR\SPURS_STD". The file name has to be specified without<br />

the *.xml file extension.<br />

The numeric suffix at SENSe is irrelevant.<br />

This command is an event and therefore has no RST value and no query.<br />

Example: "LIST:STAN:DEL 'custom_standard'"<br />

'deletes the file called custom_standard.xml<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

[SENSe:]LIST:STANdard:SAVE <br />

This command stores the current spurious emission measurement configuration to a file on the<br />

R&S <strong>FSQ</strong> hard disk.<br />

The file format is *.xml. The default path is "D:\R_S\INSTR\SPURS_STD".<br />

The numeric suffix at SENSe is irrelevant.<br />

This command is an event and therefore has no RST value and no query.<br />

Example: "LIST:STAN:SAVE 'custom_standard'"<br />

'saves the configuration to a file called custom_standard.xml<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />