R&S <strong>FSQ</strong> Instrument Functions<br />

4.7.2.2 Normalization<br />

Tracking Generator – Option R&S FSU-B9<br />

NORMALIZE The NORMALIZE softkey switches the normalization on or off. The softkey is only<br />

available if the memory contains a correction trace.<br />

REF VALUE<br />

POSITION<br />

It is possible to shift the relative reference point within the grid using the REF VALUE<br />

POSITION softkey. Thus, the trace can be shifted from the upper border of the grid<br />

to the vertical center of the grid:<br />

Fig. 4.21 Normalized display<br />

In SPLIT SCREEN operation, the normalization is switched on in the currently active<br />

window. Different types of normalization can be active in the two windows.<br />

Normalization is aborted when the NETWORK operating mode is quit.<br />

Remote command: CORR ON<br />

The REF VALUE POSITION softkey marks a reference position in the active window<br />

at which the normalization result (calculated difference with a reference trace) is displayed.<br />

If no reference line is active, the softkey switches on a reference line and activates<br />

the input of its position. The line can be moved within the grid boundaries.<br />

The reference line is switched off by pressing the softkey again.<br />

The function of the reference line is explained in the section “Calibration Mechanism”<br />

on page 4.229.<br />

Operating Manual 1313.9681.12 - 01 4.225