R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

CALCulate Subsystem<br />

"DET RMS"<br />

'Sets the RMS detector<br />

"TRIG:SOUR VID;LEV:VID 50 PCT"<br />

'Selects the trigger source VIDeo and sets the level of the video trigger source to<br />

50 PCT<br />

"SWE:TIME 50ms"<br />

'Sets the sweep time to 50 ms<br />

"INIT;*WAI"<br />

'Starts the measurement with synchronization<br />

"CALC:MARK:FUNC:MSUM? 50US,450US,576.9US,8"<br />

'Queries 8 bursts with an offset of 50 µs, a test time of 450 µs and a period of<br />

576.9 µs<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A-T<br />

CALCulate:MARKer:FUNCtion:NDBDown <br />

This command defines the level spacing of the two delta markers to the right and left of marker 1 in the<br />

selected measurement window. Marker 1 is always used as the reference marker. The numeric suffix<br />

is irrelevant for this command.<br />

The temporary markers T1 and T2 are positioned by n dB below the active reference marker. The<br />

frequency spacing of these markers can be queried with CALCulate:MARKer:FUNCtion:<br />

NDBDown:RESult?.<br />

If a negative value is entered than the markers are placed n dB above the active reference marker.<br />

This is then a n dB up function which can be used for notch filter measurements:<br />

Example: "CALC:MARK:FUNC:NDBD 3dB"<br />

'Sets the level spacing in screen A to 3 dB.<br />

Characteristics: *RST value: 6dB<br />

SCPI: device-specific<br />

Mode: A<br />

CALCulate:MARKer:FUNCtion:NDBDown:FREQuency?<br />

This command queries the two frequencies of the N-dB-down marker in the selected measurement<br />

window. The numeric suffix is irrelevant for this command. The two frequency values are<br />

separated by comma and output in ascending order.<br />

A complete sweep with synchronization to sweep end must be performed between switching on the<br />

function and querying the measured value to obtain a valid query result. This is only possible in singlesweep<br />

mode.<br />

Example: "INIT:CONT OFF"<br />

'Switches to single-sweep mode.<br />

"CALC:MARK:FUNC:NDBD ON"<br />

'Switches on the n-dB-down function in screen A.<br />

"INIT;*WAI"<br />

'Starts a sweep and waits for the end.<br />

6.58 Operating Manual 1313.9681.12 - 01