R&S <strong>FSQ</strong> Error Messages<br />

Error code Error text in the case of queue poll<br />

Error explanation<br />

9.3 Device-Specific Messages<br />

Table 9-1 Device-specific messages<br />

Device-Specific Messages<br />

-420 Query UNTERMINATED<br />

The query is incomplete.<br />

Example: The instrument is addressed as a talker and receives incomplete data.<br />

-430 Query DEADLOCKED<br />

The query cannot be processed.<br />

Example: The input and output buffers are full, the instrument cannot continue<br />

operation.<br />

-440 Query UNTERMINATED after indefinite response<br />

A query is in the same command line after a query which requests an indefinite<br />

response.<br />

Error code Error text in the case of queue poll<br />

Error explanation<br />

1036 MS: The correction table based amplifier gain exceeds the amplifier range for<br />

CALAMP1 and CALAMP2 on IF board<br />

This message is output when the setting range of the calibration amplifier is not<br />

sufficient for the required correction. The error occurs only with modules which are not<br />

correctly adjusted or defective.<br />

1052 Frontend LO is Unlocked<br />

This message is output when the phase regulation of the local oscillator fails in the RF<br />

frontend.<br />

1060 Trigger-Block Gate Delay Error- gate length < Gate Delay<br />

This message is output when the gate signal length is not sufficient for the pull-in delay<br />

with a predefined gate delay.<br />

2022 OPTIONS.INI invalid<br />

This message is output when an error has been recognized in the file OPTIONS.INI<br />

which contains the clearing codes for retrofitable firmware applications. If this file is not<br />

correctly recognized, all firmware applications are blocked for this instrument.<br />

2028 Hardcopy not possible during measurement sequence<br />

This message is output when a printout is started during scan sequences that cannot<br />

be interrupted. Such sequences are for example:<br />

• Recording the system error correction data (calibration)<br />

• Instrument selftest<br />

In such cases a synchronization to the end of the scan sequence should be performed<br />

prior to starting the printout.<br />

2033 Printer Not Available<br />

This message is output when the selected printer is not included in the list of available<br />

output devices. A possible cause is that the required printer driver is missing or<br />

incorrectly installed.<br />

2034 CPU Temperature is too high<br />

This message is output when the temperature of the processor exceeds 70 °C.<br />

9.8 Operating Manual 1313.9681.12 - 01