R&S <strong>FSQ</strong> Instrument Functions<br />

C/N /<br />

C/NO<br />

C/N /<br />

C/NO<br />

CHANNEL<br />

BANDWIDTH<br />

ADJUST<br />

SETTINGS<br />

<strong>Analyzer</strong> Mode<br />

The C/N C/NO softkey opens the submenu for configuring the carrier/noise ratio<br />

measurement.<br />

C/N / C/NO<br />

CHANNEL BANDWIDTH<br />

ADJUST SETTINGS<br />

The user can choose between measurement without (C/N) and measurement with<br />

reference to the bandwidth (C/No). In addition, it is possible to select the bandwidth<br />

of the channel and to adapt the span.<br />

The measurements are only available in the frequency domain (span >0).<br />

The C/N and C/NO softkeys enable and disable the measurement of the carrier/<br />

noise ratio, the C/No measurement also being referred to a 1 Hz bandwidth.<br />

The maximum value of the current trace is determined when the function is activated<br />

and is marked by means of the REFERENCE FIXED marker.<br />

The measurement is performed on the trace where MARKER 1 is located. To measure<br />

another trace, MARKER 1 has to be shifted to the trace in question using the<br />

SELECT TRACE softkey in the MARKER menu.<br />

If no marker is active, MARKER 1 is activated when the function is switched on.<br />

Remote command: CALC:MARK:FUNC:POW:SEL CN<br />

CALC:MARK:FUNC:POW:RES? CN<br />

CALC:MARK:FUNC:POW:SEL CN0<br />

CALC:MARK:FUNC:POW:RES? CN0<br />

CALC:MARK:FUNC:POW OFF<br />

The CHANNEL BANDWIDTH softkey opens a window for selecting the measurement<br />

channel bandwidth.<br />

The default setting is 14 kHz.<br />

The specified channel bandwidth allows the optimal setting of the measurement<br />

parameters of the R&S <strong>FSQ</strong> using ADJUST SETTINGS.<br />

Remote command: SENS:POW:ACH:BWID 14kHz<br />

The ADJUST SETTINGS softkey adapts the span to the channel bandwidth<br />

selected.<br />

For the carrier/noise ratio measurement, the span is set to:<br />

4 × channel bandwidth + measurement margin<br />

The adjustment is performed once; if necessary, the setting can be changed later<br />

on.<br />

Remote command: SENS:POW:ACH:PRES CN | CN0<br />