R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

CALCulate:DELTamarker:MAXimum:RIGHt<br />

CALCulate Subsystem<br />

This command positions the specified delta marker to the next smaller maximum value to the right of<br />

the current value (i.e. ascending X values). The corresponding delta marker is activated first, if<br />

necessary.<br />

Example: "CALC2:DELT:MAX:RIGH"<br />

'Sets delta marker 1 in screen B to the next smaller maximum value to the right of<br />

the current value.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is an event and therefore has no *RST value and no query.<br />

CALCulate:DELTamarker:MINimum:LEFT<br />

This command positions the specified delta marker to the next higher minimum value to the left of the<br />

current value (i.e. descending X values). The corresponding delta marker will be activated first, if<br />

necessary.<br />

Example: "CALC:DELT:MIN:LEFT"<br />

'Sets delta marker 1 in screen A to the next higher minimum to the left of the current<br />

value.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is an event and therefore has no *RST value and no query.<br />

CALCulate:DELTamarker:MINimum:NEXT<br />

This command positions the specified delta marker to the next higher minimum value of the trace. The<br />

corresponding delta marker will be activated first, if necessary.<br />

Example: "CALC1:DELT2:MIN:NEXT"<br />

'Sets delta marker 2 in screen A to the next higher minimum value.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is an event and therefore has no *RST value and no query.<br />

CALCulate:DELTamarker:MINimum[:PEAK]<br />

This command positions the specified delta marker to the current minimum value of the trace. The<br />

corresponding delta marker will be activated first, if necessary.<br />

Example: "CALC2:DELT3:MIN"<br />

'Sets delta marker 3 in screen B to the minimum value of the associated trace.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is an event and therefore has no *RST value and no query.<br />

6.18 Operating Manual 1313.9681.12 - 01