18.12.2012 Views

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

[SENSe:]LIST:RANGe:INPut:ATTenuation:AUTO ON | OFF<br />

This command switches the autoranging of a range in the spurious measurement on or off.<br />

Example: "LIST:RANG2:INP:ATT:AUTO ON"<br />

'Activates autorange for range 2<br />

Characteristics: *RST value: ON<br />

SCPI: device-specific<br />

Mode: A<br />

[SENSe:]LIST:RANGe:INPut:GAIN:STATe ON | OFF<br />

This command switches the preamplifier of a range in the spurious measurement on or off.<br />

Example: "LIST:RANG2:INP:GAIN:STAT ON"<br />

'Activates the preamplifier for range 2<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

[SENSe:]LIST:RANGe:LIMit -200dB to +200dB<br />

SENSe Subsystem<br />

This command defines the limit lines for the spurious emission measurement list evaluation (for details<br />

on this measurement see “LIST EVALUATION” on page 4.131). For each of the 20 ranges, you can<br />

specify a limit line via the numeric suffix at RANGe. The numeric suffix at SENSe is<br />

irrelevant.<br />

Example: "LIST:RANG5:LIM -40"<br />

’Limit line in range 5 is –40 dB.<br />

Characteristics: *RST value: -13dB<br />

SCPI: device-specific<br />

Mode: A<br />

[SENSe:]LIST:RANGe:LIMit:STATe ON | OFF<br />

This command activates/deactivates the limit checking for the spurious emission measurement list<br />

evaluation (for details on this measurement see “LIST EVALUATION” on page 4.131).<br />

The numeric suffix at SENSe and the numeric suffix at RANGe1...20> are irrelevant.<br />

Example: "LIST:RANG5:LIM -40"<br />

’Limit line in range 5 is –40 dB.<br />

"LIST:RANG7:LIM -20"<br />

’Limit line in range 7 is–20 dB.<br />

"LIST:RANG:LIM:STAT ON"<br />

’Activates the limit checking in all ranges.<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

Operating Manual 1313.9681.12 - 01 6.183

Hooray! Your file is uploaded and ready to be published.