R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

Example: "CALC:MARK:MAX:AUTO ON"<br />

Activates the auto search function for marker 1.<br />

Characteristics: *RST value: OFF<br />

SCPI: device-specific<br />

Mode: A<br />

CALCulate:MARKer:MAXimum:LEFT<br />

CALCulate Subsystem<br />

This command positions the specified marker to the next smaller maximum value to the left of the<br />

current value (i.e. in descending X values) on the trace.<br />

Example: "CALC:MARK2:MAX:LEFT"<br />

'Positions marker 2 in screen A to the next lower maximum value to the left of the<br />

current value.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is an event and therefore has no *RST value and no query.<br />

CALCulate:MARKer:MAXimum:NEXT<br />

This command positions the specified marker to the next smaller maximum value of the corresponding<br />

trace.<br />

Example: "CALC:MARK2:MAX:NEXT"<br />

'Positions marker 2 in screen A to the next 'lower maximum value.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

If no next smaller maximum value is found on the trace (level spacing to adjacent<br />

values < peak excursion), an execution error (error code: -200) is produced.<br />

If no next smaller maximum value is found on the trace (level spacing to adjacent<br />

values < peak excursion), an execution error (error code: -200) is produced.<br />

This command is an event and therefore has no *RST value and no query.<br />

CALCulate:MARKer:MAXimum[:PEAK]<br />

This command positions the specified marker to the current maximum value of the corresponding<br />

trace. The corresponding marker is activated first or switched to the marker mode.<br />

If no maximum value is found on the trace (level spacing to adjacent values < peak<br />

excursion), an execution error (error code: -200) is produced.<br />

Operating Manual 1313.9681.12 - 01 6.47