R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.11 INITiate Subsystem<br />

INITiate Subsystem<br />

The INITiate subsystem controls the init-measurement function in the selected measurement window.<br />

The measurement windows are assigned to INITiate1 (screen A) and INITiate2 (screen B).<br />

INITiate:CONMeas<br />

This command continues a stopped measurement at the current position in single-sweep mode. The<br />

function is useful especially for trace functions MAXHold, MINHold and AVERage if the previous<br />

results are not to be cleared with Sweep Count > 0 or Average Count > 0 on restarting the<br />

measurement (INIT:IMMediate resets the previous results on restarting the measurement).<br />

The single-sweep mode is automatically switched on. Synchronization to the end of the indicated<br />

number of measurements can then be performed with the command *OPC, *OPC? or *WAI. In the<br />

continuous-sweep mode, synchronization to the sweep end is not possible since the overall<br />

measurement "never" ends.<br />

Example: "INIT2:CONT OFF"<br />

'Switches to single-sweep mode.<br />

"DISP:WIND:TRAC:MODE AVER<br />

'Switches on trace averaging.<br />

"SWE:COUN 20"<br />

Setting the sweep counter to 20 sweeps.<br />

"INIT2;*WAI"<br />

'Starts the measurement and waits for the end of the 20 sweeps.<br />

"INIT2:CONM;*WAI"<br />

'Continues the measurement (next 20 sequences) and waits for the end.<br />

Characteristics: *RST value: -<br />

SCPI: device-specific<br />

Mode: A<br />

This command is an event and therefore has no *RST value and no query.<br />

INITiate:CONTinuous ON | OFF<br />

This command determines whether the trigger system is continuously initiated (continuous) or<br />

performs single measurements (single).<br />

In analyzer mode, this setting refers to the sweep sequence (switching between continuous/single<br />

sweep).<br />

Example: "INIT:CONT OFF"<br />

'Switches the sequence in screen B to single sweep.<br />

"INIT:CONT ON"<br />

'Switches the sequence to continuous sweep.<br />

Characteristics: *RST value: ON<br />

SCPI: conform<br />

Mode: A<br />

Operating Manual 1313.9681.12 - 01 6.121