18.12.2012 Views

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

R&S FSQ Signal Analyzer - Rohde & Schwarz

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

R&S <strong>FSQ</strong> Instrument Functions<br />

SWEEP<br />

CONTINUOUS<br />

SWEEP<br />

SINGLE<br />

SWEEP<br />

CONTINUE<br />

SGL SWEEP<br />

4.5.5 Sweep Settings – SWEEP Key<br />

<strong>Analyzer</strong> Mode<br />

The SWEEP key serves for configuring the sweep mode and opens the SWEEP<br />

softkey menu. In split-screen mode, the entries made are valid for the active window<br />

only.<br />

The CONTINUOUS SWEEP, SINGLE SWEEP and SGL SWEEP DISP OFF softkeys<br />

are mutually exclusive selection keys.<br />

CONTINUOUS SWEEP<br />

SINGLE SWEEP<br />

CONTINUE SGL SWEEP<br />

SWEEPTIME MANUAL<br />

SWEEPTIME AUTO<br />

SWEEP COUNT<br />

SWEEP POINTS<br />

SGL SWEEP DISP OFF<br />

The CONTINUOUS SWEEP softkey activates the continuous sweep mode, which<br />

means that the sweep takes place continuously according to the trigger mode set.<br />

When working in the split-screen mode and with different settings in the two windows,<br />

screen A is swept first, followed by screen B. When the softkey is pressed,<br />

the sweep is restarted.<br />

CONTINUOUS SWEEP is the default setting of R&S <strong>FSQ</strong>.<br />

Remote command: INIT:CONT ON<br />

The SINGLE SWEEP softkey starts n sweeps after triggering. The number of<br />

sweeps is determined by the SWEEP COUNT softkey.<br />

When working in the split-screen mode, the frequency ranges of the two windows<br />

are swept one after the other.<br />

If a trace is swept using TRACE AVERAGE or MAXHOLD, the value set via the<br />

SWEEP COUNT softkey determines the number of sweeps. If 0 has been entered,<br />

one sweep is performed.<br />

Remote command: INIT:CONT OFF<br />

The CONTINUE SGL SWEEP softkey repeats the number of sweeps set under<br />

SWEEP COUNT. You have to measure with a sweep count >1, else the R&S <strong>FSQ</strong><br />

deletes the trace before resuming the sweep.<br />

This is particularly of interest when using the functions TRACE AVERAGE and<br />

MAXHOLD, if previously recorded measurement results are to be taken into consideration<br />

for averaging / maximum search.<br />

If SGL SWEEP DISP OFF is active, the screen is switched off also during repeated<br />

sweeps.<br />

Remote command: INIT:CONM<br />

Operating Manual 1313.9681.12 - 01 4.31

Hooray! Your file is uploaded and ready to be published.

Saved successfully!