R&S <strong>FSQ</strong> Remote Control – Description of Commands<br />

6.5.5.2 CALCulate:LIMit:ACPower Subsystem<br />

CALCulate Subsystem<br />

The CALCulate:LIMit:ACPower subsystem defines the limit check for adjacent channel power measurement.<br />

CALCulate:LIMit:ACPower:ACHannel:ABSolute -200DBM to 200DBM, -200 to<br />

200DBM<br />

This command defines the absolute limit value for the lower/upper adjacent channel during adjacent<br />

channel power measurement (Adjacent Channel Power) in the selected measurement window.<br />

It should be noted that the absolute limit value has no effect on the limit check as soon as it is below<br />

the relative limit value defined with CALCulate:LIMit:ACPower:ACHannel:RELative. This<br />

mechanism allows automatic checking of the absolute basic values of adjacent channel power as<br />

defined in mobile radio standards.<br />

The numeric suffixes in LIMIt are irrelevant for this command.<br />

Parameter: The first value is the limit for the lower and the upper adjacent channel. The second<br />

limit value is ignored but must be indicated for reasons of compatibility with the FSE<br />

family.<br />

Example: "CALC:LIM:ACP:ACH:ABS -35DBM, -35DBM"<br />

'Sets the absolute limit value in screen A for the power in the lower and upper<br />

adjacent channel to -35 dBm.<br />

Characteristics: *RST value: -200DBM<br />

SCPI: device-specific<br />

Mode: A<br />

CALCulate:LIMit:ACPower:ACHannel:ABSolute:STATe ON | OFF<br />

This command activates the limit check for the adjacent channel when adjacent channel power<br />

measurement (Adjacent Channel Power) is performed. Before the command, the limit check for the<br />

channel/adjacent channel measurement must be globally switched on using CALC:LIM:ACP ON.<br />

The result can be queried with CALCulate:LIMit:ACPower:ACHannel:RESult?. It should be<br />

noted that a complete measurement must be performed between switching on the limit check and the<br />

result query, since otherwise no valid results are available.<br />

The numeric suffixes in LIMIt are irrelevant for this command.<br />

Example: "CALC:LIM:ACP:ACH 30DB, 30DB"<br />

'Sets the relative limit value in screen A for the power in the lower and upper<br />

adjacent channel to 30 dB below the channel power.<br />

"CALC:LIM:ACP:ACH:ABS -35DBM, -35DBM"<br />

'Sets the absolute limit value in screen A for the power in the lower and upper<br />

adjacent channel to -35 dBm.<br />

"CALC:LIM:ACP ON"<br />

'Switches on globally the limit check for the channel/adjacent channel<br />

measurement in screen A.<br />

"CALC:LIM:ACP:ACH:REL:STAT ON"<br />

'Switches on the check of the relative limit values for adjacent channels in<br />

screen A.<br />

"CALC:LIM:ACP:ACH:ABS:STAT ON"<br />

'Switches on the check of absolute limit values for the adjacent channels in<br />

screen A.<br />

6.30 Operating Manual 1313.9681.12 - 01