R&S <strong>FSQ</strong> Instrument Functions<br />

REF POINT<br />

LEVEL<br />

REF POINT LVL<br />

OFFSET<br />

REF POINT<br />

FREQUENCY<br />

<strong>Analyzer</strong> Mode<br />

The REF POINT LEVEL softkey activates an entry box for the input of a reference<br />

level other than the reference marker level. The function is identical to that of the<br />

softkey with the same name in the MARKER menu (MKR).<br />

Remote command: CALC:DELT1:FUNC:FIX:RPO:Y -10dB<br />

The REF POINT LVL OFFSET softkey activates an entry box for the input of an<br />

additional level offset for the phase noise calculation.<br />

This level offset is set to 0 dB on when the REFERENCE FIXED or PHASE NOISE<br />

function is enabled.<br />

Remote command: CALC:DELT:FUNC:FIX:RPO:Y:OFFS 10dB<br />

The REF POINT FREQUENCY softkey activates an entry box for the manual input<br />

of a reference frequency for the REFERENCE FIXED or PHASE NOISE function.<br />

Remote command: CALC:DELT1:FUNC:FIX:RPO:X 10.7MHz<br />

PEAK SEARCH The PEAK SEARCH sets the reference point level for delta marker 2 in the selected<br />

measurement window to the peak of the selected trace.<br />

AUTO PEAK<br />

SEARCH<br />

Remote command: CALC:DELT:FUNC:FIX:RPO:MAX<br />

The AUTO PEAK SEARCH softkey activates an automatic peak search for the reference<br />

fixed marker 1 at the end of each particular sweep.<br />

This function may be used for tracking of a drifting source during a phase noise<br />

measurement. The delta marker 2, which shows the phase noise measurement<br />

result, keeps the delta frequency value. Therefore the phase noise measurement<br />

leads to reliable results in a certain offset although the source is drifting. Only if the<br />

marker 2 is reaching the border of the span, the delta marker value is adjusted to be<br />

within the span. In these cases, choose a larger span.<br />

Remote command: CALC:DELT:FUNC:PNO:AUTO ON | OFF<br />

Measurement example<br />

The phase noise of a CW signal at 100 MHz with 0 dBm level is to be measured at<br />

800 kHz from the carrier<br />

[PRESET] The R&S <strong>FSQ</strong> is set to the default setting.<br />

[CENTER: 100 MHz] The center frequency is set to 100 MHz.<br />

[SPAN: 2 MHz] The span is set to 2 MHz.<br />

[AMPT: 0 dBm] The reference level is set to 0 dBm.<br />

[MKR FCTN] MARKER 1 is switched on and positioned at the maximum<br />

of the displayed trace.<br />

[PHASE NOISE: 800 kHz] The phase noise measurement is switched on. The<br />

delta marker is positioned on the main marker and the<br />

measured phase noise value is displayed in the marker<br />

info field. The sample detector is used and the video<br />

bandwidth is set to 3 × RBW. When the phase noise<br />

measurement function is enabled, the entry of the delta<br />

marker frequency is activated. It can be entered<br />

directly.<br />

4.66 Operating Manual 1313.9681.12 - 01